Call

Blog

Dental X-ray Room Requirements in Ontario

Panoramic dental X-ray image showing the full upper and lower jaw

Short answer

In Ontario, no one may install a dental X-ray machine until the Director of X-ray Safety has issued written approval, and that approval covers the shielding of the room. The plans include a floor plan at a scale of not less than one to fifty, submitted in duplicate, showing the position and limits of rotation of the chair and the position of the patient’s head. Doors into the room need self-closing devices, and warning signs where the public can reach them.

Dental X-ray room requirements in Ontario come from the Healing Arts Radiation Protection Act, known as the HARP Act, and its regulation, the X-ray Safety Code (Reg. 543). They apply whether the unit sits in an operatory or in a separate imaging room, and they apply to new offices, renovations and equipment replacements alike.

This guide explains what the rules require before, during and after construction, and how the room itself is built. It is written for dentists planning work in Toronto and the GTA, and it reflects how Odima Construction coordinates X-ray rooms within our wider medical and healthcare construction work.

Who Regulates Dental X-ray Rooms in Ontario

Two levels of government are involved. The federal Radiation Emitting Devices Act deals with the sale, lease and import of devices: no person shall sell, lease or import into Canada a radiation emitting device that does not comply with the prescribed standards. Health Canada notes that provincial and territorial legislation or professional bodies may set requirements for safely installing and using these devices.

In Ontario, installation and use fall under the HARP Act, which the Minister administers through the Director of X-ray Safety. The RCDSO states that an X-ray Inspection Service (xRIS) inspector at the Ministry may visit any dental facility, and that written approval from xRIS is required before installing a new dental X-ray machine. The RCDSO sets professional duties for dentists, while approval of the installation stays with the Director.

Overhead view of a dental chair, delivery unit and handpieces on a wood-look floor, the chair position an X-ray plan must show

Dental X-ray Room Requirements Before Installation

Under section 3 of the HARP Act, no person shall install an X-ray machine unless the Director has issued written approval for the installation. The Act says a reference to installation includes the shielding of the area where the machine is installed, so the room construction is part of what is approved. Dental machines are covered; there is no dental exemption.

The application includes the plans, specifications and information set out in Reg. 543. These include:

  • The owner’s name and an identifier for the room.
  • The manufacturer and model number of the X-ray machine, the anticipated maximum workload, the maximum tube voltage and the maximum tube current.
  • The materials of the room’s boundaries and the occupancy of adjacent spaces.
  • A floor plan drawn to a scale of not less than one to fifty, submitted in duplicate.
  • For a dental facility, the position and limits of rotation of the chair and the position of the head of the person being irradiated.

Because the floor plan must show the chair and head positions, the operatory layout has to be settled before the submission. Our guide to planning a dental office floor plan explains how operatory layout and imaging rooms fit together.

Shielding Limits and Materials

Reg. 543 requires every installation to be shielded with a primary protective barrier and a secondary protective barrier, so that no X-ray worker receives more than 1 millisievert per week and no other person, other than a patient, receives more than 0.1 millisievert per week. The regulation names the standards for designing that shielding: Health Canada Safety Code 20A, Appendix II (1999), or NCRP Report No. 147 (2005).

The Ontario Building Code adds a parallel sentence: every installation of an X-ray machine or X-ray equipment in a building shall be shielded to protect any person who could be exposed to radiation. Its explanatory note says the barriers should be designed and installed to comply with the HARP Act.

In practice, lead is commonly used, often as lead-lined drywall, and the shielding designer sets where it goes and how high it runs based on the equipment, workload, layout and what is on the other side of each wall. Once approved, the machine is installed only in accordance with the plans the approval was based on.

Doors, Signs and Room Size

Reg. 543 says doors into an X-ray room shall be fitted with self-closing devices and, where the doors are accessible to the public, shall have warning signs prominently displayed on them. Door hardware, frames and any glazing are therefore chosen with the shielding design in mind.

There is no standard room dimension in these rules. Room size follows from the chair, the delivery system, the X-ray unit’s mounting position and its clearance drawings, and the limits of chair rotation shown on the plan. The RCDSO also says the owner must keep an up-to-date approved floor plan, so later changes to the room should be planned with that drawing in hand.

Steel stud wall framing with drywall edges and a wood panel, the stage at which shielded wall layers are installed in an X-ray room

Registration, the RPO and Testing After Installation

The HARP Act says the owner shall not operate an X-ray machine unless the machine, its location and the owner’s name and business address are registered with the Director. Since a December 2025 amendment, the owner must also give a current email address and notify the Director in writing of changes to business or email addresses.

The RCDSO states that in a dental facility, only a dentist can be the radiation protection officer. Under Reg. 543, the radiation protection officer must send the Director, within sixty days of the installation of a new X-ray machine, test results showing compliance with the federal Radiation Emitting Devices Act. The RCDSO X-ray guidance for dentists also covers thyroid shields and the photographic quality assurance program required in every dental office.

Renovations, Replacements and Dental CT

A change to an approved installation needs the Director’s written approval for the change, and the RCDSO says approval is required when replacing or changing a dental installation as well as for a new machine. Moving an operatory, opening a wall in an X-ray room or swapping a unit therefore starts with the approval process. Our article on renovating a dental office in Ontario explains how this fits into a wider project.

Dental CT has extra rules. The HARP Act requires a CT scanner to be in a facility designated by the Minister, and the RCDSO states that starting August 15, 2023, dental facilities replacing an aging CT scanner must submit a request to xRIS and receive approval before installing and operating the new machine. The RCDSO notes this does not affect its facility permit requirements.

How Odima Builds X-ray Rooms

Odima coordinates the trades, building systems and owner-supplied equipment interfaces for the room: framing, shielded wall layers, doors and hardware, blocking for the X-ray arm, power and data. We build to the approved plans and shielding design, and we raise any field change before it is made so the practice can seek approval for it. Our dental office construction service includes this coordination, and our dental clinic build-out and renovation team can review your plans. To start, share your floor plan and equipment list.

Frequently Asked Questions

Do I need approval before installing a dental X-ray machine in Ontario?

Yes. The HARP Act says no person shall install an X-ray machine unless the Director of X-ray Safety has issued written approval, and installation includes the shielding of the room. The RCDSO confirms dentists need written approval from the X-ray Inspection Service before installing a new dental X-ray machine, and when replacing or changing an installation.

What are the standard dimensions for an X-ray room?

Ontario’s rules do not set a standard dimension for a dental X-ray room. The size follows from the chair, the delivery system, the X-ray unit’s mounting and clearance drawings, and the limits of chair rotation that the approval plan must show. The designer sizes the room around the equipment you choose.

Who regulates radiation in Canada?

It is shared. The federal Radiation Emitting Devices Act covers the sale, lease and import of devices such as dental X-ray units. Provinces set rules for installing and using them; in Ontario that is the HARP Act and the X-ray Safety Code, administered through the Director of X-ray Safety at the Ministry.

Who is ultimately responsible for the radiation safety of a dental facility?

The practice. The owner of the X-ray machine must register the machine, its location and the owner’s name and business address with the Director before operating it, and must keep an up-to-date approved floor plan. In a dental facility, only a dentist can be the radiation protection officer.

What is the best material to shield against radiation?

Shielding is designed to the standards Reg. 543 names, Health Canada Safety Code 20A Appendix II or NCRP Report No. 147, so the right material and thickness depend on the equipment, workload and adjacent spaces. Lead is commonly used, often as lead-lined drywall, and the shielding designer specifies where it goes.

What do the doors of a dental X-ray room need?

Reg. 543 requires doors into an X-ray room to be fitted with self-closing devices. Where the doors are accessible to the public, warning signs must be prominently displayed on them. Door frames, hardware and any glazing are selected together with the approved shielding design so the barrier stays continuous.

Does a dental CT scanner have different requirements?

Yes. The HARP Act requires a CT scanner to be in a facility designated by the Minister. The RCDSO states that since August 15, 2023, dental facilities replacing an aging CT scanner must request and receive xRIS approval before installing and operating the new machine, and that this does not change the College’s facility permit requirements.

Plan Your Dental X-ray Room With Odima

Send us your floor plan, your imaging equipment list and your target dates, and we will coordinate the room construction with your approved plans and shielding design. Odima Construction works with dental practices across Toronto and the GTA.

Request a Quote Call 416-777-6666

Planning a Clinic Build-Out?

Odima Construction builds and renovates medical, dental, physiotherapy, and rehab clinics across Toronto and the GTA. Send us your floor plan and target dates, and we will arrange a time to meet.

Request a Quote 416-777-6666
Call 416-777-6666 Book a Consultation